Acute Hemodynamic Efficacy of a 32-ml Subcutaneous
Counterpulsation Device in a Calf Model of Diminished Cardiac
Function
Steven C. Koenig*,†, Kenneth N. Litwak‡, Guruprasad A. Giridharan*, George M.
Pantalos*,†, Robert D. Dowling†, Sumanth D. Prabhu§, Mark S. Slaughter¶, Michael A.
Sobieski¶, and Paul A. Spence||
* Department of Bioengineering, Cardiovascular Innovation Institute, University of Louisville,
Louisville Kentucky
NIH-PA Author Manuscript
†
Department of Surgery, Louisville, Kentucky
‡
Cleveland Clinic, Cleveland, OH
§
Department of Medicine, Institute of Molecular Cardiology, University of Louisville, Louisville,
Kentucky
¶
Advocate Christ Medical Center, Oak Lawn, Illinois
||
SCR, Inc., Louisville, Kentucky
Abstract
NIH-PA Author Manuscript
The acute hemodynamic efficacy of an implantable counter-pulsation device (CPD) was evaluated.
The CPD is a valveless single port, 32-ml stroke volume blood chamber designed to be connected
to the human axillary artery using a simple surface surgical procedure. Blood is drawn into the pump
during systole and ejected during diastole. The acute hemodynamic effects of the 32-ml CPD were
compared to a standard clinical 40-ml intra-aortic balloon pump (IABP) in calves (80 kg, n = 10).
The calves were treated by a single oral dose of Monensin to produce a model of diminished cardiac
function (DCF). The CPD and IABP produced similar increases in cardiac output (6% CPD vs. 5%
IABP, p > 0.5) and reduction in left ventricular external work (14% CPD vs. 13% IABP, p > 0.5)
compared to DCF (p < 0.05). However, the ratio of diastolic coronary artery flow to left ventricular
external work increase from DCF baseline (p < 0.05) was greater with the CPD compared to the
IABP (15% vs. 4%, p < 0.05). The CPD also produced a greater reduction in left ventricular
myocardial oxygen consumption from DCF baseline (p < 0.05) compared to the IABP (13% vs. 9%,
p < 0.05) despite each device providing similar improvements in cardiac output. There was no early
indication of hemolysis, thrombus formation, or vascular injury. The CPD provides hemodynamic
efficacy equivalent to an IABP and may become a therapeutic option for patients who may benefit
from prolonged counterpulsation.
The intra-aortic balloon pump (IABP) is the most widely used short-term cardiac assist device
with over 160,000 balloons implanted each year world-wide. There is also clinical evidence
to suggest that counterpulsation therapy can be effective in patients with chronic congestive
heart failure (CHF). The IABP has been shown to be effective in patients with poor left
ventricular function undergoing coronary artery bypass surgery.1 However, the location of an
Reprint Requests: Steven C. Koenig, PhD, Cardiovascular Innovation Institute, 302 East Muhammad Ali Blvd, room 408, Departments
of Biomedical Engineering and Surgery, University of Louisville, Louisville, KY 40202. sckoen01@gwise.louisville.edu.
Koenig et al.
Page 2
NIH-PA Author Manuscript
IABP catheter (thoracic aorta) and biocompatibility issues limit the application of IABP
therapy to short durations of hours to <14 days. When IABP support is employed for prolonged
periods of more than 20 days, the frequency of vascular complications, infection, and bleeding
is high.2,3 Furthermore, the IABP immobilizes the patient.
It is not surprising that a number of companies are now developing chronic heart assist devices
to permit patients to be treated with long-term counterpulsation therapy that avoids catheter
placement in the groin and allows patients to be mobile [left ventricular assist devices (LVAD)
Technologies (CardioPlus4,5), Sunshine Heart (C-Pulse6), Cardiak (akpulsor), and SCR7,8].
These devices are being developed to offer the advantages of long-term counterpulsation with
an implant procedure that is less invasive than current LVAD.
SCR Counterpulsation Device Therapy
NIH-PA Author Manuscript
A simple counterpulsation device (CPD) that can be implanted and explanted with a minimally
invasive surface procedure is currently under development (SCR, Louisville, KY). The core
component of the SCR device is a 32-ml stroke volume polyurethane-lined blood chamber,
(Figure 1) designed to fit comfortably in a subcutaneous pocket, (7 cm × 7 cm × 5 cm) on the
right anterior chest contralateral to an implantable cardioverter defibrillator (ICD) device
implantation site. The pumping chamber is connected to the systemic circulation by a short
ePTFE graft anastomosed to the right axillary artery. A percutaneous drive line runs from the
chamber to an external pneumatic driver. During systole, the drive console evacuates air from
the chamber thus removing blood from the arterial circulation reducing cardiac work and afterload. During diastole, the console ejects blood from the chamber into the circulation providing
diastolic augmentation and improving coronary perfusion. The chamber and graft have been
specifically developed for continuous washing and nominal shear stress. The shape of the CPD
chamber is designed to wash continuously by creating a constant but shifting circular vortex
of blood to minimize risk of blood stasis and thrombosis. The CPD graft connects seamlessly
to the chamber without an inlet or outlet valve, also reducing the risk of thrombus formation.
The bi-directional flow enhances washing inside the graft compared to unidirectional cannulae
used in other blood pumps. The CPD graft is also supported with reinforcement rings and the
distance between the chamber and axillary artery is minimized to reduce the risk of graft
kinking.
Methods
All animals received humane care in accordance with the Guide for Care and Use of Laboratory
Animals (NIH publication 86–23, revised 1996) and the guidelines determined by the
Institutional Animal Care and Use Committee of the University of Louisville.
NIH-PA Author Manuscript
Experimental Design
The acute hemodynamic efficacy of a 32-ml CPD chamber (SCR, Louisville, KY) and a
standard commercially-available 40-ml IABP (Datascope, Fairfield, NJ) were investigated in
a large animal model (n = 10) of diminished cardiac function (DCF) created with a toxin
(Monensin). The CPD device was connected to the carotid artery (6–8 mm diameter) with a
10 mm ePTFE graft and the IABP was placed in the descending thoracic aorta. The carotid
artery diameter of the calf and distance from CPD to aorta are equivalent in size of human
axillary artery and distance to aorta. Pressure, flow, and volume waveforms were recorded and
myocardial oxygen consumption was calculated during 1) baseline (device off), 2) 1:1 support,
and 3) 1:2 support for the 32-ml CPD and 40-ml IABP in the same animal. Baseline
measurements were repeated between test conditions to ensure DCF baseline values were
maintained at a comparable steady state. Baselines and support modes for each device were
maintained for 15 minutes each to ensure a steady-state was reached while minimizing risk of
ASAIO J. Author manuscript; available in PMC 2010 April 5.
Koenig et al.
Page 3
NIH-PA Author Manuscript
deterioration of the animal preparation. Data were collected for 30 seconds at the end of each
steady-state test condition. The test order of devices (IABP and CPD) and support modes (1:1
and 1:2) were randomized to eliminate experimental order bias. Cardiovascular parameters
were derived and left ventricular pressure-volume (PV) loops constructed from recorded
waveforms to assess the hemodynamic efficacy of each device as defined by changes in left
ventricular workload, coronary perfusion, and myocardial oxygen consumption. A blood
sample was drawn for all test conditions to determine plasma free hemoglobin (PFHb) as an
early indication of device associated hemolysis.
Surgical Procedures (DCF Model)
NIH-PA Author Manuscript
Ten healthy Jersey calves (male, 80–90 kg) were used in this study. After initial physical
examination and a 5–7 day quarantine period, animals were preanesthetized with Atropine (30
mg) and then anesthetized with isoflurane (3%–5%) via nose cone. After intubation, anesthesia
was maintained with isoflurane (1%–3%) mixed with oxygen and room air. The left lateral
chest was prepared for sterile surgery. A thoracotomy was performed over the 3rd and 4th ribs.
Catheters (Millar Instruments, Houston, TX) were temporarily placed in the left atrium (5F
single-tip model MPR-500) and across aortic valve (6F dual-tip conductance model SPR-562)
for simultaneous measurement of aortic pressure and left ventricular pressure and volume. A
flow probe (Transonics, Ithaca, NY) was temporarily placed around the pulmonary artery.
Pressure and flow waveforms were recorded to assess normal baseline cardiac parameters. The
catheters and flow probe were then removed. Fluid-filled arterial and venous pressure lines
were implanted in the left carotid artery and jugular vein to monitor cardiac function during
the recovery period. The baseline cardiac output in these animals is approximately 10 L/min.
In order to reduce cardiac output for testing clinically sized devices, a cardiotoxin was
administered. After a 1-week recovery period, calves were given a single oral dose of Rumensin
80 equivalent to 20 mg/kg (containing 80 mg/lb of Monensin: Elanco, Indianapolis, IN) via
orogastric tube to create the DCF model. Monensin is a common cattle herd diet additive that
has been shown to be a cardiotoxic producing extensive diffuse cardiomyocyte death in calves
when used at these doses. Previous investigations have shown that this model can reduce
cardiac output by 15%–25%, increase left atrial pressure by 67%, and increase left ventricular
end-diastolic pressure by 143% in calves.9 Higher doses of Monesin produce greater reduction
in cardiac output, but at the risk of an unstable model and high mortality.
Surgical Procedures (Acute Efficacy Study)
NIH-PA Author Manuscript
Three weeks after administering the Monensin dose, the nonsurvival acute hemodynamic study
was initiated. Anesthesia, surgery (thoracotomy), and instrumentation procedures were
repeated as described above with the addition of a flow probe placed around the left main
coronary artery. Blood sampling catheters were placed in the aorta via the femoral artery and
in the coronary sinus via the hemiazygous vein. A 10-mm ePTFE vascular graft (Atrium,
Hudson, NH) was anastomosed to the right carotid artery and attached to the 32-ml CPD. A
40-ml IABP catheter (Datascope, Fairfield, NJ) was inserted into the descending thoracic aorta
immediately downstream to the aortic arch vessels. The 32-ml CPD and 40-ml IABP were
triggered from a surface ECG and the timing was adjusted using the aortic pressure trace. IABP
timing was adjusted to inflate the balloon at the dicrotic notch and deflate the balloon
immediately before ventricular systole to maximize the reduction of aortic end-diastolic
pressure. CPD timing was adjusted to fill during left ventricular ejection and eject immediately
after the dicrotic notch. The CPD filling occurs over a longer time period to prevent suction of
the carotid artery that can occur from filling the CPD chamber too quickly. Animals were
anticoagulated with heparin (3 mg/kg i.v.) and ACT was maintained above 200 seconds. Blood
was sampled every 30 minutes to check ACT. Central venous pressure (CVP) was measured
with a fluid-filled pressure line and monitored for observation by our anesthesia technician and
surgeon throughout each experiment. The surgical team carefully monitored CVP to ensure it
ASAIO J. Author manuscript; available in PMC 2010 April 5.
Koenig et al.
Page 4
was maintained at approximately10–12 mm Hg for each baseline pre-, post-, and in-between
IABP and CPD test conditions.
NIH-PA Author Manuscript
Instrumentation and Data Reduction
All transducers were pre- and postcalibrated against known physical standards to ensure
measurement accuracy. Hemodynamic data were signal conditioned and A/D converted for
digital analysis using a Good Laboratory Practices (GLP) compliant data acquisition system.
10,11 Signal conditioning was accomplished using transducer amplifiers (Ectron, San Diego,
CA), transit-time flowmeter (Transonics, Ithaca, NY), and a volume conductance unit
(Leycom, Sigma V, Netherlands). These data were low-pass filtered at 60 Hz, analog-todigitally converted (AT-MIO-16E-10 and LabVIEW, National Instruments) at a sampling rate
of 400 Hz, and stored on a personal computer for postprocessing and analysis.
NIH-PA Author Manuscript
The measured pressure and flow waveform recordings were used to calculate heart rate, stroke
volume, cardiac output, CPD stroke volume, mean aortic and mean aortic diastolic pressures,
left ventricular peak systolic and end-diastolic pressures, mean left atrial pressure, mean
diastolic coronary artery flow, mean coronary artery flow, and left ventricular external work.
These parameters were calculated on a beat-to-beat basis, for each 30 second data set, using a
Hemodynamic Evaluation and Assessment Research Tool (HEART) program12 developed in
Matlab (MathWorks, Natick, MA). All analyzed beats in each data set (approximately 40–50
beats) were averaged to obtain a single representative mean value for each calculated
parameter. Left ventricular PV loops were constructed by plotting left ventricular pressure
against left ventricular volume with each loop representing one complete cardiac cycle. These
PV loops were used to calculate left ventricular external work and evaluate changes in left
ventricular end-diastolic and end-systolic pressures and volumes. Statistical analyses were
conducted using t test for paired samples and assuming equal variances. Statistical analysis
was performed to test for changes in the cardiac parameters between the normal and the DCF
state. Statistical analysis was also performed to test for efficacy of the devices compared to the
DCF state and to compare devices.
Left Ventricular Myocardial Oxygen Consumption
NIH-PA Author Manuscript
Paired blood samples were drawn from the aorta and coronary sinus to determine the oxygen
content of the blood entering and exiting the left ventricular myocardium at the end of each
15-minute test condition while left main coronary artery blood flow (CAF, ml/min) was
recorded. The blood samples were processed through a blood gas analyzer (IRMASL Blood
Analysis System, Diametics Medical, Inc., St. Paul, MN) to measure the % saturation of
hemoglobin with oxygen (%O2 Sat) and hemoglobin content [(Hb)gm/dl]. The oxygen content
(O2) was then calculated for each blood sample by,
The left ventricular myocardial oxygen consumption (LV O2 cons.) was calculated as,
ASAIO J. Author manuscript; available in PMC 2010 April 5.
Koenig et al.
Page 5
[O2]a = arterial oxygen content, [O2]cs = coronary sinus oxygen content, and LV mass [g] =
weight of LV at necropsy.
NIH-PA Author Manuscript
Blood Hemoglobin and Necropsy
Venous blood samples were drawn to measure Total Hemoglobin (Tot Hb) and Plasma Free
Hemoglobin (PFHb) for each experimental test condition and device trial. At the completion
of the experiment, the CPD chamber and cannula were removed and visually inspected for
gross evidence of thrombus, fractures, and other defects. Peripheral organs were removed and
examined for gross evidence of thrombus related infarcts.
Results
Monensin-Induced DCF Animal Model
The oral dose of Monensin resulted in a 17% reduction in cardiac output (reduced from 10.3
to 8.5 L/min), a 9% reduction in left ventricular peak systolic pressure, an 11% reduction in
mean arterial pressure, a 120% increase in left atrial pressure, and a 133% increase in left
ventricular end diastolic pressure compared to normal baseline (Table 1).
Acute Efficacy Study
NIH-PA Author Manuscript
Example aortic pressure traces for unassisted support (baseline) and 40-ml IABP (top) and 32ml CPD (bottom) in 1:1 assist mode are shown side-by-side in Figure 2. As indicated by the
arrows, the peak systolic aortic pressures were reduced by a similar amount with both devices.
Example aortic pressure traces for the CPD (dotted line) and IABP (solid line) are also shown
overlaid in Figure 3. As indicated by the arrows, the IABP lowered end diastolic pressure more
than the CPD.
Statistically significant (p < 0.05) hemodynamic improvements were observed with both the
32-ml CPD and 40-ml IABP compared to DCF baseline values (Table 2). The CPD and IABP
increased cardiac output by 6%. The mean diastolic coronary artery flow was augmented by
14% with the CPD and 7% with the IABP. Left ventricular peak systolic pressure was reduced
by 7% with the CPD and 6% with the IABP. There were no statistically significant differences
between the 32-ml CPD and 40-ml IABP (p > 0.5), except for the aortic end-diastolic pressure,
for which the IABP produced a greater reduction (p < 0.05).
NIH-PA Author Manuscript
Superimposed coronary flow waveforms for the IABP (solid line) and the CPD (dotted line)
are shown in Figure 4. Coronary flow was reversed prior to systole with the IABP (shaded
area), but did not occur with the CPD. Thus, the CPD provided a significantly higher
augmentation in coronary flow than the IABP (14% increase with CPD vs. 7% increase with
IABP).
A sample left ventricular pressure volume (PV) loop is shown in Figure 5. The 40-ml IABP
and 32-ml CPD both reduced left ventricular volume and systolic pressures from DCF baseline
as evidenced by the leftward and downward shifts in the PV loops. The left ventricular external
work (equivalent to the area inside the PV loop) was reduced by both the 32-ml CPD (14%)
and the 40-ml IABP (13%) compared to DCF baseline (Figure 6).
The myocardial perfusion to work balance was also considerably better with the 32-ml CPD
as evidenced by the significantly increased diastolic coronary artery flow to left ventricular
external work ratio (increased 15% by CPD vs. 4% IABP), as shown in Figure 7. In addition,
the 32-ml CPD resulted in a greater reduction in the left ventricular myocardial oxygen
consumption (decreased 13% by CPD vs. 9% IABP) despite the fact that both devices provided
similar improvements in cardiac output.
ASAIO J. Author manuscript; available in PMC 2010 April 5.
Koenig et al.
Page 6
Blood Hemoglobin and Necropsy
NIH-PA Author Manuscript
Plasma free hemoglobin levels remained below 8 mg/dl at all measured time points. A necropsy
was performed at the end of each animal study and included examination of all major abdominal
and thoracic organs, and the brain. There were no signs of damage or thromboembolism in any
organ. The CPD was examined carefully for signs of thrombus. The CPD chamber was entirely
clear of thrombi for all experiments.
Discussion
Congestive heart failure, one of the largest unsolved problems in cardiac care today, affects
more that five million patients in the United States alone. Despite pharmacological therapy,
many patients deteriorate to the point where mechanical circulatory support would be clinically
beneficial. Unfortunately, despite many years of research and millions of dollars of
expenditure, only about 3,000 patients receive chronic mechanical circulatory support each
year. Most current mechanical circulatory support requires invasive sternotomy and
cannulation of the heart and aorta. Complications are frequent and include device failure,
thromboembolism, stroke, infection, and bleeding. The level of invasiveness of the surgery
required for implantation, the difficulty in the management of the patients, and the frequency
of complications contribute to the low number of patients receiving mechanical heart assist
devices.
NIH-PA Author Manuscript
Globally, the most common assist device for heart failure is the IABP, which is used clinically
for limited time periods of several hours to several days. Each year more than 160,000 patients
worldwide receive IABP therapy, often for cardiogenic shock or severe left ventricular
dysfunction following acute myocardial infarction or cardiotomy.13,14 Many thousands of lives
are saved each year due to IABP therapy (65% survival).15 Intra-aortic balloon pump support
is simpler than current LVAD therapies and clinicians have developed a great deal of
confidence in the therapeutic effectiveness of intra-aortic balloon counterpulsation. The IABP
requires no direct cannulation of the heart leading to easier implantation and explantation.
Furthermore, control algorithms to fill and empty the balloon are considerably less complex
than to control an LVAD and simplify patient management. Short-term counterpulsation
therapy, although unquestionably less powerful than LVAD therapy, does offer considerable
practical clinical advantages. With over 30 years of use, clinicians trust counterpulsation
therapy in their sickest patients.
NIH-PA Author Manuscript
Counterpulsation therapy is achieved by rapidly inflating the IABP immediately after aortic
valve closure (dicrotic notch) and rapidly deflating the balloon just before the onset of systole.
The rapid inflation of the balloon increases the aortic diastolic pressure by 30%–70%
improving end-organ perfusion and coronary perfusion.13,14 The rapid deflation of the IABP
reduces the ejection pressure of the native ventricle, reducing after load and left ventricular
external work.13,16,17 The peak systolic pressure is reduced by up to 15%,13,17 whereas the
end diastolic pressure is decreased by up to 30%.1,17 Cardiac output and stroke volume have
been found to increase by up to 20%,16,18 while no changes have been observed.14,16,18 Intraaortic balloon pump therapy also decreases the heart rate by 10%14 and may reduce the left
ventricular end diastolic volume by 10%–15%.19 Intra-aortic balloon pump improves the
cardiac energy balance by increasing the oxygen supply to the myocardium through increased
coronary perfusion of 0%–100%20,21 and by reducing myocardial oxygen consumption
through a decrease in after-load and left ventricular external work.14 Intra-aortic balloon pump
also has been reported to augment cerebral, renal, mesenteric, and pulmonary blood flows.
22–24 Further, increases in urine output and lactate use, decrease in lactic acidosis and
enhancement of venous oxygen saturation have also been observed in patients supported with
IABP.14
ASAIO J. Author manuscript; available in PMC 2010 April 5.
Koenig et al.
Page 7
NIH-PA Author Manuscript
Counterpulsation can also be performed with a blood chamber connected directly to the aorta.
Nanas et al. showed that a 30-ml blood chamber was more effective in reducing aortic systolic
and end-diastolic pressures and myocardial oxygen consumption while increasing coronary
flow when directly compared to a 40-ml IABP.5 In our laboratory, computer simulation and
feasibility experiments have suggested that a 32-ml blood chamber connected to the axillary
artery can be as effective as a 40-ml IABP in the descending aorta.7,8 The results of this study
demonstrate that the 32-ml CPD anastamosed to a peripheral vessel (carotid artery) produced
hemodynamic efficacy equivalent to a standard clinical 40-ml IABP. Importantly, in this study,
the 32-ml CPD provided greater diastolic myocardial perfusion per unit work (supply) with
lower myocardial oxygen consumption (demand) compared to the IABP. This study also
demonstrated that there were no early indicators of hemolysis, thrombus formation, or vascular
injury associated with the 32-ml CPD.
NIH-PA Author Manuscript
Counterpulsation therapy is most effective when the balloon or blood chamber volumes are
comparable to the ventricular stroke volume and when the vasculature is stiffer.25–29 Calves
weighing 80 –90 kg were selected for this study because the size of their carotid arteries (6
mm) and the distance from the anastomotic site to the aortic valve (14 cm) approximated the
clinical situation that will be encountered with a 32-ml CPD anastomosed to the axillary artery
(8 mm diameter and 12 cm to the aortic valve). In these acute experiments, a reproducible
patent end-to-side anastomosis was achieved with a 10 mm graft. Calves have higher cardiac
outputs and stroke volumes (even with cardiac dysfunction produced by the DCF model) and
higher arterial compliance than humans. Therefore, the improvement in cardiac performance
was less pronounced in this study than would be expected in the clinical application of
counterpulsation. Importantly, despite the limitations of the calf model, the improvements in
cardiac function produced by the 32-ml CPD were equivalent to those produced by the 40-ml
IABP.
NIH-PA Author Manuscript
There are important distinctions in the method of delivery and timing of counterpulsation
therapy between the IABP and the CPD. The IABP displaces blood volume within the aorta
while the CPD chamber shuttles blood in and out of the aorta. The ejection and filling phases
of the CPD are longer than the balloon inflation and deflation time periods of the IABP. The
faster balloon deflation is necessary to prevent the obstruction of the aorta during ventricular
systole. However, this rapid deflation, leads to reverse coronary flow (Figure 3). The IABP
rapidly inflates and deflates during diastole within the same assisted beat. In contrast, the CPD
fills during systole and ejects during diastole. In this study, these differences in timing resulted
in a lower aortic end diastolic pressure and diastolic coronary flow with the 40-ml IABP
compared with the CPD (Figures 3 and 4). Ultimately, the IABP and CPD both produced
comparable after load reduction compared to unassisted condition when operating in 1:1 or 1:2
mode. The myocardial perfusion and perfusion to work balance was superior with the CPD
compared to the IABP.
Counterpulsation with IABP has not gained widespread acceptance as a long-term therapy to
treat heart failure, because of the high risk of complication and lack of patient mobility. There
are reports that counterpulsation can have a sustained effect when used for prolonged periods
of support.25 Recently, Jeevanandam reported that patients who received an aortic CPD
(CardioPlus, LVAD Technologies) demonstrated considerable improvements in exercise
capacity.4,5 Unfortunately, implantation of this device required a thoracotomy and resulted in
a high level of morbidity and mortality for these patients.
SCR (Louisville, KY) is developing a chronic implantable CPD to treat HF patients who may
be responsive to a partial form of cardiac assist. It is intended to be a low cost device that can
be implanted and explanted without major surgical trauma or extensive postoperative care. The
device is implanted by a surface procedure in a pacemaker pocket on the right side and
ASAIO J. Author manuscript; available in PMC 2010 April 5.
Koenig et al.
Page 8
NIH-PA Author Manuscript
connected to the axillary artery. The size and shape of the SCR device are designed to provide
a good anatomical fit in humans while closely matching the expected levels of support needed
for effective counterpulsation therapy support in HF patients. Chronic testing of the CPD is
on-going. The CPD has successfully supported two calves chronically for 30 days without
adverse events or device failures.
Conclusion
This acute study demonstrated that the 32-ml CPD and the 40-ml IABP produced an equivalent
and statistically significant improvement in cardiac performance compared to the DCF state.
The 32-ml CPD did not cause any acute hemolysis, thrombus formation, or vascular injury.
Acknowledgments
Supported by grants from the National Institutes of Health (NIH) Heart Lung and Blood Institute (HLBI) Small
Business Innovative Research (SBIR) program (grant number 1R43HL083586-01) and the Kentucky Science and
Engineering Foundation (KSEF) program (grant number KSTC-184-512–07-011).
References
NIH-PA Author Manuscript
NIH-PA Author Manuscript
1. Cochran RP, Starkey TD, Panos AL, Kunzelman KS. Ambulatory intraaortic balloon pump use as
bridge to heart transplant. Ann Thorac Surg 2002;74:746–751. discussion 751–752. [PubMed:
12238834]
2. Freed PS, Wasfie T, Zado B, et al. Intraaortic balloon pumping for prolonged circulatory support. Am
J Cardiol 1988;61:554–557. [PubMed: 3344678]
3. Manord JD, Garrard CL, Mehra MR, et al. Implications for the vascular surgeon with prolonged (3 to
89 days) intraaortic balloon pump counterpulsation. J Vasc Surg 1997;26:511–516. [PubMed:
9308597]
4. Jeevanandam V, Jayakar D, Anderson AS, et al. Circulatory assistance with a permanent implantable
IABP: Initial human experience. Circulation 2002;24:I183–I188. [PubMed: 12354730]
5. Terrovitis JV, Charitos CE, Tsolakis EJ, et al. Superior performance of a paraaortic counterpulsation
device compared to the intraaortic balloon pump. World J Surg 2003;27:1311–1316. [PubMed:
14586569]
6. Legget ME, Peters WS, Milsom FP, et al. Extra-aortic balloon counterpulsation: An intraoperative
feasibility study. Circulation 2005;112:I26–I31. [PubMed: 16159827]
7. Giridharan GA, Koenig SC, Pantalos GM, et al. Predicted hemodynamic benefits of counterpulsation
therapy using a superficial surgical approach. ASAIO J 2006;52:39–46. [PubMed: 16436889]
8. Koenig SC, Litwak KN, Pantalos GM, et al. Development and early testing of a simple subcutaneous
circulatory support device. ASAIO J 2006;52:362–367. [PubMed: 16883113]
9. Litwak KN, McMahan A, Lott KA, et al. Monensin toxicosis in the bovine calf: A large animal model
of cardiac dysfunction. Comp Med 2005;44:45–49.
10. Drew, GA.; Koenig, SC. Biomedical Patient Monitoring, Data Acquisition, and Playback with
LabVIEW®. In: Olansen, JB.; Rosow, E., editors. Virtual Bio-Instrumentation: Biomedical, Clinical,
and Healthcare Applications in LabVIEW®. Prentice Hall; 2002. p. 180-186.
11. Koenig SC, Woolard C, Drew GD, et al. Integrated data acquisition system for medical device testing
and physiology research in compliance with good laboratory practices. Biomed Instrum Technol
2004;38:229–240. [PubMed: 15174367]
12. Schroeder MJ, Perrrault B, Ewert DL, Koenig SC. HEART: An automated beat-to-beat cardiovascular
analysis package using Matlab™. Comput Biol Med 2004;34:371–388. [PubMed: 15145710]
13. Nanas JN, Moulopoulos SD. Counterpulsation: Historical background, technical improvements,
hemodynamic and metabolic effects. Cardiology 1994;84:156–167. [PubMed: 8205565]
14. Papaioannou GT, Stefanadis C. Basic principles of the intraaortic balloon pump and mechanisms
affecting its performance. ASAIO J 2005;51:296–300. [PubMed: 15968962]
ASAIO J. Author manuscript; available in PMC 2010 April 5.
Koenig et al.
Page 9
NIH-PA Author Manuscript
NIH-PA Author Manuscript
15. Torchiana DF, Hirsch G, Buckley MJ, et al. Intraaortic balloon pumping for cardiac support: trends
in practice and outcome, 1968 to 1995. J Thor Cardiovasc Surg 1997;13:758–764.
16. Scheidt S, Wilner G, Mueller H, et al. Intra-aortic balloon counterpulsation in cardiogenic shock.
Report of a cooperative clinical trial. N Engl J Med 1973;288:979–984. [PubMed: 4696253]
17. Bolooki, H., editor. Clinical Application of the Intra-aortic Balloon Pump. 3. New York: Futura; 1998.
18. Dunkman WB, Leinbach RC, Buckley MJ, et al. Clinical and hemodynamic results of intra-aortic
balloon pumping and surgery for cardiogenic shock. Circulation 1972;46:465–477. [PubMed:
4561118]
19. Weiss AT, Engel S, Gotsman CJ, et al. Regional and global left ventricular function during intraaortic balloon counterpulsation in patients with acute myocardial infarction shock. Am Heart J
1984;108:249–254. [PubMed: 6464962]
20. Kern MJ, Aguirre FV, Tatineni S, et al. Enhanced coronary blood flow velocity during intra-aortic
balloon counterpulsation in critically ill patients. J Am Coll Cardiol 1993;21:359–368. [PubMed:
8425999]
21. Ishihara M, Sato H, Tateishi H, et al. Effects of intraaortic balloon pumping on coronary
hemodynamics after coronary angioplasty in patients after an acute myocardial infarction. Am Heart
J 1992;124:1133–1138. [PubMed: 1442477]
22. Bhayana JN, Scott SM, Sethi GK, Takaro T. Effects of intra-aortic balloon pumping on organ
perfusions in cardiogenic shock. J Surg Res 1979;26:108–113. [PubMed: 423560]
23. Sugita Y, Emoto H, Morita K, et al. The effect of intraaortic balloon pumping (IABP) on pulmonary
circulation. ASAIO Trans 1985;31:389–394.
24. Hilberman M, Derby GC, Spencer RJ, Stinson EB. Effect of the intra-aortic balloon pump upon
postoperative renal function in man. Crit Care Med 1981;9:85–89. [PubMed: 7460589]
25. Sun Y. Modeling the dynamic interaction between left ventricle and intra-aortic balloon pump. Am
J Physiol 1991;261:H1300–H311. [PubMed: 1928410]
26. Papaioannou TG, Mathioulakis D, Nanas J, et al. Arterial compliance is a main variable determining
the effectiveness of intra-aortic balloon counterpulsation: Quantitative data from an in-vitro study.
Med Eng Physics 2002;24:279–284.
27. Lin C-Y, Galysh FT, Ho KJ, Patel AS. Response to single-segment intraaortic balloon pumping as
related to aortic compliance. Ann Thorac Surg 1972;13:468–476. [PubMed: 5022424]
28. Papaioannou TG, Lekakis J, Dagre A, et al. Arterial compliance is an independent factor that predicts
acute hemodynamic performance of intra-aortic balloon counterpulsation. Int J Artif Organs
2001;24:478–483. [PubMed: 11510920]
29. Feola M, Haiderer O, Kennedy JG. Intra-aortic balloon pumping (IABP) at different levels of
experimental acute left ventricular failure. Chest 1971;59:68–76. [PubMed: 5540966]
NIH-PA Author Manuscript
ASAIO J. Author manuscript; available in PMC 2010 April 5.
Koenig et al.
Page 10
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Figure 1.
Photograph of implantable components of the SCR counterpulsation device (CPD) system: 1)
chamber, 2) connector, and 3) pneumatic driveline.
NIH-PA Author Manuscript
ASAIO J. Author manuscript; available in PMC 2010 April 5.
Koenig et al.
Page 11
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Figure 2.
Example aortic pressure traces for unassisted support (baseline) and 40-ml IABP (top) and 32ml CPD (bottom) in 1:1 assist mode. As indicated by the arrows, the peak systolic aortic
pressures were reduced by a similar amount with both devices.
NIH-PA Author Manuscript
ASAIO J. Author manuscript; available in PMC 2010 April 5.
Koenig et al.
Page 12
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Figure 3.
NIH-PA Author Manuscript
Example aortic pressure traces for 40-ml IABP (solid) and 32-ml CPD (dotted) in 1:1 assist
mode. As indicated by the arrows, the 40-ml IABP resulted in a lower end diastolic pressure
than the CPD.
ASAIO J. Author manuscript; available in PMC 2010 April 5.
Koenig et al.
Page 13
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Figure 4.
NIH-PA Author Manuscript
Example coronary flow waveform traces for 40-ml IABP (solid) and 32-ml CPD (dotted) in
1:1 assist mode. Coronary flow was reversed prior to systole with the IABP (shaded area) but
coronary flow reversal did not occur with the CPD.
ASAIO J. Author manuscript; available in PMC 2010 April 5.
Koenig et al.
Page 14
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Figure 5.
NIH-PA Author Manuscript
A sample left ventricular (LV) pressure volume (PV) loop. The 40-ml IABP and 32-ml CPD
both reduced left ventricular volume and systolic pressures from DCF baseline as evidenced
by the leftward and downward shifts in the PV loops.
ASAIO J. Author manuscript; available in PMC 2010 April 5.
Koenig et al.
Page 15
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Figure 6.
The left ventricular external work (LVEW) was reduced by both the 32-ml CPD and 40-ml
IABP. [*Vs. DCF baseline (p < 0.05)].
NIH-PA Author Manuscript
ASAIO J. Author manuscript; available in PMC 2010 April 5.
Koenig et al.
Page 16
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Figure 7.
The 32-ml CPD provided a greater increase in mean diastolic coronary artery flow (CAFd) per
unit left ventricular external work (LVEW) and greater reduction in left ventricular myocardial
oxygen consumption (LV O2). These data demonstrate the 32-ml CPD delivers more
myocardial perfusion (supply) while also reducing cardiac work (demand). [*vs. DCF baseline
(p < 0.05) and #CPD Vs. IABP (p < 0.05)].
NIH-PA Author Manuscript
ASAIO J. Author manuscript; available in PMC 2010 April 5.
NIH-PA Author Manuscript
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Table 1
HR (bpm)
SV (ml)
CO (L/min)
LVPpksys (mm Hg)
MAP (mm Hg)
LAP (mm Hg)
LVPed (mm Hg)
Normal
90 ± 5
118 ± 10
10.3 ± 0.5
116 ± 3
91 ± 4
5±2
6±2
DCF
77 ± 3*
110 ± 7
8.5 ± 0.6*
106 ± 3*
81 ± 3*
11 ± 2*
14 ± 3*
Koenig et al.
The DCF Model Produced a Significant Reduction in Cardiac Output (CO) While Left Atrial Pressure (LAP) More Than Double
Values are given as mean ± SD.
*
Normal vs. DCF (p < 0.05).
ASAIO J. Author manuscript; available in PMC 2010 April 5.
HR, heart rate; SV, stroke volume; LVPpksys, left ventricular peak systolic pressure; MAP, mean aortic pressure; LAP, left atrial pressure; LVPed, left ventricular end diastolic pressure at normal (before oral
dose of Monensin); DCF, diminished cardiac function.
Page 17
NIH-PA Author Manuscript
NIH-PA Author Manuscript
NIH-PA Author Manuscript
Table 2
Device
Mode
HR (bpm)
SV (ml)
CO (L/min)
CAFdia (ml/min)
MAP (mm Hg)
AoPed (mm Hg)
77 ±3
110 ±7
8.5 ±0.6
106 ±3
279 ±51
81 ±3
74 ±5
1:1
79 ±4
113 ±9
9.0 ±0.9*
99 ±3*
317 ±52*
86 ±4*
68 ±3*
1:2
79 ±4
113 ±8
9.3 ±0.6*
100 ±3*
297 ±49*
85 ±3*
68 ±3*
1:1
79 ±3
113 ±7
9.0 ±0.5*
96 ±3*
297 ±70*
85 ±3*
60 ±3*†
1:2
80 ±4
111 ±4
8.9 ±0.6*
95 ±2*
294 ±64*
81 ±3*
60 ±3*†
DCF baseline
32-ml CPD
40-ml IABP
LVPpksys (mm Hg)
Koenig et al.
The 32-ml CPD and 40-ml IABP Both Produced a Significant Hemodynamic Improvement Compared to DCF Baseline
ASAIO J. Author manuscript; available in PMC 2010 April 5.
Values are given as mean [±] SD.
*
Vs. DCF baseline (p < 0.05).
†
IABP vs. CPD (p < 0.05).
HR, heart rate; SV, stroke volume; CO, cardiac output; LVPpksys, left ventricular peak systolic pressure; CAFdia, mean diastolic coronary artery flow; MAP, mean aortic pressure; AoPed, aortic end-diastolic
pressure at DCF, diminished cardiac function baseline, 1:1 and 1:2 modes.