This article appeared in a journal published by Elsevier. The attached
copy is furnished to the author for internal non-commercial research
and education use, including for instruction at the authors institution
and sharing with colleagues.
Other uses, including reproduction and distribution, or selling or
licensing copies, or posting to personal, institutional or third party
websites are prohibited.
In most cases authors are permitted to post their version of the
article (e.g. in Word or Tex form) to their personal website or
institutional repository. Authors requiring further information
regarding Elsevier’s archiving and manuscript policies are
encouraged to visit:
http://www.elsevier.com/copyright
Author's personal copy
Radiation Physics and Chemistry 80 (2011) 917–922
Contents lists available at ScienceDirect
Radiation Physics and Chemistry
journal homepage: www.elsevier.com/locate/radphyschem
Thermal neutron fluence in a treatment room with a Varian linear accelerator
at a medical university hospital
Wen-Shan Liu a, Sheng-Pin Changlai b, Lung-Kwang Pan c, Hsien-Chun Tseng a, Chien-Yi Chen d,e,n
a
School of Medicine, Chung Shan Medical University and Department of Radiation Oncology, Chung Shan Medical University Hospital, Taiwan, Republic of China
Department of Nuclear Medicine, Lin Shin Hospital, Taichung 402, Taiwan, Republic of China
c
Department of Medical Imaging and Radiological Sciences, Central Taiwan University of Science and Technology, Takun 40609, Taiwan, Republic of China
d
Department of Medical Image, Chung Shan Medical University Hospital, Taiwan, Republic of China
e
School of Medical Imaging and Radiological Sciences, Chung Shan Medical University, Taichung 402, Taiwan, Republic of China
b
a r t i c l e i n f o
abstract
Article history:
Received 14 February 2008
Accepted 31 March 2011
Available online 8 April 2011
The indium foil activation technique has been employed to measure thermal neutron fluences (Fth)
among various locations in the treatment room with a 20 20 cm2 field size and a 15 and 10 MV X-ray
beam. Spatial Fth are visualized using colored three-dimensional graphical representations; intensities
are up to (1.97 7 0.13) 105 and (1.46 7 0.13) 104 n cm 2/Gy-X at isocenter, respectively. The Fth is
found to increase with the X-ray energy of the LINAC and decreases as it moves away from the beam
center. However, thermal neutron exposure is not assessed in routine dosimetry planning and radiation
assessment of patients since neutron dose contributes o 1% of the given therapy dose. However, unlike
the accelerated beam limited within the gantry window, photoneutrons are widely spread in the
treatment room. Distributions of Fth were measured in water phantom irradiated with 15 MV X-ray
beams. The shielding effect of the maze was also evaluated. The experimentally estimated Fth along the
maze distance was fitted explicate and the tenth-value layer (TVL) was calculated and discussed. Use of
a 10 cm-thick polyethylene door placed at the maze was suitable for radiation shielding.
& 2011 Elsevier Ltd. All rights reserved.
Keywords:
Indium foils activation technique
Thermal neutron fluences
Water phantom
Shielding effect
1. Introduction
High-energy electron accelerators, such as the electron linear
accelerator (LINAC) are utilized routinely to generate photon beams
for cancer therapy (Kamino, 1998). However, when high-energy
accelerators are operated roughly at 10 MV or higher, a significant
number of neutrons are generated through photonuclear reactions
via (e, n) and (x, n), such that high-energy bremsstrahlung react with
the target, dynamic wedge and multi-leaves resulting in a mixed
radiation field in the radiotherapy area (Konefal et al., 2005; Lennox,
2001; Lin et al., 2001; Kase et al., 1998; AAPM, 1986). With growing
interest in the use of high-energy photon beams at Chung Shan
Medical University Hospital (CSMUH), detailed measurements of
thermal neutron fluences (Fth) are required. Fth have been extensively investigated using calculations and measurement techniques
for LINACs with various energies. Table 1 lists the neutron fluences
in the 10–25 MV range (Barquero et al., 2005; Konefal et al., 2005;
Lin et al., 2001; Paredes et al., 1999; Palta et al., 1984; Gur et al.,
1978; Deye and Young, 1977; McGinley et al., 1976; Uwamino et al.,
1986). The foil activation technique has been employed for
n
Corresponding author at: School of Medical Imaging and Radiological Sciences,
Chung Shan Medical University, Taiwan, Republic of China.
Tel.: þ886 4 24730022 x 17220; fax: þ886 4 247217210.
E-mail address: ccy@csmu.edu.tw (C.-Y. Chen).
0969-806X/$ - see front matter & 2011 Elsevier Ltd. All rights reserved.
doi:10.1016/j.radphyschem.2011.03.022
three-dimensional distribution measurements of Fth. Neutrons are,
however, scattered in the human body (Comsan, 1996). The neutron
field in the body differs from that in the primary field. Distributions
of Fth depth were measured in a water phantom irradiated with a
15 MV X-ray beam. The problem of adequate shielding from maze
and neutron scattering into the maze may require additional attention. The tenth-value layer (TVL) of Fth was expressed as an
exponential function to the calculated points in the maze.
2. Materials and methods
2.1. Varian CLINAC 21EX
A large number of LINACs around the world are currently in use
for radiation therapy (Baker and Thomas, 2001; Lennox, 2001;
Kamino, 1998). However, only limited studies are focused on the
distributions of Fth inside treatment rooms, water phantoms and
mazes of medical LINACs. A detailed survey of Fth in treatment
rooms, water phantoms and mazes is important for radiotherapy
patients. The LINAC (Varian CLINAC 21EX; Palo Alto, CA, USA) at
CSMUH is designed for Intensity Modulated Radiation Therapy
(IMRT), which is achieved by superposing different MLC segments.
Approximately 8500–9000 patients with various pathologies, most of
them with nasopharyngeal cancer in the head and neck have been
treated at CSMUH since 1999. LINAC generates dual photon energies
Author's personal copy
918
W.-S. Liu et al. / Radiation Physics and Chemistry 80 (2011) 917–922
Table 1
Neutron fluence measurements of this study and other investigations.
Accelerator
Energy (MV)
SSD (cm)
X-ray filed size (cm2)
Method
Neutron fluence
n cm 2/Gy-X
Ref.
Varian CLINAC 21EX
Varian CLINAC 21EX
Varian 2300
Siemens
Varian CLINAC 2300
Varian CLINAC 2300
Primus Siemens
Siemens
Siemens Primus
Varian CLINAC 2100 C
Mevatron 77
Philips SL/75-20
Varian CLINAC
Varian CLINAC-18
Allis-Chalmers Betatron
Siemens
Microtron, MM22
15
10
22
21
20
18
18
15
15
18
18
18
18
10
25
18
21
100
100
20 20
20 20
In foil
In foil
100
10 10
In foil
40 40
P2O5
CR 39, MCNP
Au foil
In foil
5.1 cm Rh
In foil
In foil, BF3
BF3
In foil
1.97 105
1.46 104
1.10 104
2.00 104
1.51 106
7.00 103
1.20 104
2.90 105
1.51 106
1.07 105
2.30 105
1.4 106
1.4 104
6.80 104
3.80 105
4.60 105
3.2 105
This study
This study
Konefal et al., 2005
Konefal et al., 2005
Konefal et al., 2005
Konefal et al., 2005
Konefal et al., 2005
Konefal et al., 2005
Lin et al., 2001
Paredes et al., 1999
Palta et al., 1984
Gur et al., 1978
Deye and Young, 1977
McGinley et al., 1976
McGinley et al., 1976
Barquero et al., 2005
Uwamino et al., 1986
100
100
100
100
100
15 15
10 10
25 25
10 10
10 10
100
10 10
Table 2
Systematic and random errors for the practical evaluation of photo neutron
fluence in this study.
concrete
2.2 m
concrete
2.7 m
concrete
y
2.2 m
x
Source
One standard deviation Di
Others:
Daily X-ray constancy
Daily electron output constancy
LINAC power fluctuation
o2%
o2%
o1%
Measurement:
Internal normalization of foil
Activated foil counting statistics
1.1–1.8%
6.12% 9.4%
Dtot
6.90–10.2%
Isocenter
(0,0,0)
slide door
A
250 mm Pb
Maze
B
1.0 m
S
Control
Area
Fig. 1. Plane view and measurement locations in the Varian CLINAC 21EX
treatment room at CSMUH. The 10 cm-thick PE sliding door was installed to
decrease transmitted neutron fluence.
(C-69; Wellhofer Co., Germany) calibrated by the National Ionization
Standard Laboratory, Institute of Nuclear Energy Research. The errors
associated with the derived neutron fluence were calculated as
the square root of the sum of the squares of the individual errors
Di, as listed in Table 2. The experimental uncertainty includes (1) the
dosimetry and power fluctuation of LINAC, which were less than
2% and (2) counting statistics during In foil activation analysis.
2.2. Foil activation technique
The foil activation technique is widely utilized for measuring
with accelerating voltages up to 6 MV or higher. The electrons are
accelerated and hit the gantry head target (Lennox, 2001; Kamino,
1998). Fig. 1 shows a plane view of the radiotherapy area with the
Varian CLINAC 21EX at CSMUH. The accelerator isocenter (0, 0 and 0)
is located midway between the primary wall barriers at a height of
1.3 m above the floor (Lennox, 2001; Kamino, 1998). The treatment
room has a metal slab and 2.2 m thick concrete on the left side of the
primary barriers. Right primary wall barriers are fabricated entirely
with 2.7 m thick concrete. A 250 mm thick lead slab was employed
in the construction of the left side of the primary barriers. The
isocenter is 3.55 m from the primary barrier. The treatment room is
7.9 m long by 7.1 m wide by 3.8 m high. A slide door of 10 cm
polyethylene (PE) is used to moderate the fast and intermediate
energy neutrons (Baker and Thomas, 2001). Lead of 1 cm thick is
installed at the maze end. Due to its weight, the sliding door
has a motorized opener. The maze below the treatment room is
located between the control area of right and approximately rectangular concrete of 6.8 m long 2.7 m wide 3.8 m high (Fig. 1).
The X-ray dose rate was confirmed using an ionization chamber
Fth (n cm 2/Gy-X). Thermal neutrons can be identified by selecting appropriate foil materials (Table 1; Barquero et al., 2005;
Konefal et al., 2005; Chao et al., 2001; Lin et al., 2001; Paredes
et al., 1999; Palta et al., 1984; Gur et al., 1978; Deye and Young,
1977; McGinley et al., 1976; Uwamino et al., 1986). The Fth in the
treatment room was measured using the following typical conditions with a source-surface distance (SSD) of 100 cm and 10
photon Gy for each experimental run of 2.5 min. All experiments
were conducted in triplicate. (1) Experiments were repeated with
collimator opened to 20 20 cm2 and the gantry was vertically
oriented pointing down at the floor. (2) The accelerators were
operated at 15 MV in water phantom.
The Fth in the treatment room was measured by activation of
indium foils via 115In(n, g)116m1In reaction, which has high
activation section (162 barns) for the thermal neutrons and a
suitably short half-life (T1/2 ¼54.2 min) (Konefal et al., 2005;
Chao et al., 2001; AAPM, 1986; Gur et al., 1978). An indium foil
(purity 499.999%; 25 mm long 25 mm wide 1 mm high,
Goodfellow Cambridge, Ltd., UK) with an average mass of
Author's personal copy
W.-S. Liu et al. / Radiation Physics and Chemistry 80 (2011) 917–922
919
15 MV. The water phantom base was 58 cm long 41 cm
wide 35.5 cm high; the polymethyl-methacrylate (PMMA) tank
had a volume of approximately 7100 cm3. The PMMA tank was
filled with water to a height of 30 cm (Fig. 2(b)). The water
phantom center was placed directly on the X-ray beam and
located at the isocenter (i.e., 15 cm below the water surface).
These indium foils were inserted into Plexiglas holders. Each
indium foil was placed individually exactly 1 cm apart on the x, y
and z axes. In total, 69 detection points were performed subsequently at incremental steps to map the Fth in the water
phantom. These phantom measurements can be performed at
positions of the central axis ranging in depth from 1 to 30 cm and
at two 30 cm off-axis position at 15 cm depth.
z
isocenter
x
y
2.4. The maze design of the 15 MV treatment room at CSMUH
Fig. 2. Measurements of Fth performed in the water phantom during exposure to
15 MV X-rays. Indium foil locations are indicated by circles (K); the x and y axes
correspond to locations in the treatment room for SSD ¼ 100 cm for 15 and 10 MV
X-ray exposures, respectively.
4.86 70.07 g was placed at the isocenter as a reference foil, other
identical indium foils were suspended anywhere inside the room
during experiments (Fig. 2(a); AAPM, 1986; Palta et al., 1984).
Consider a thin indium foil of mass m exposed for a period of time
ti. After the irradiation if the foil is allowed to cool for time td and
then counted for time tc using a high-purity germanium (HPGe)
detector (GC3520; Canberra Industries, Meriden, CT, USA), which
has an active volume of 145.3 cm3, 35% relative efficiency and
resolution of 2.0 keV at 1332 keV, Fth can be derived as
Fth ¼
NIn
egsK
ð1Þ
where K ¼ nð1elti Þðeltd Þð1eltc Þ=l, NIn is the number of activated nuclides of 116m1In, l is the decay constant (¼0.693/T1/2), e is
the detection efficiency; s is the microscopic cross-section, T1/2 is
the half-life of 116m1In, and g is the absolute g-ray intensity (29.2%).
n¼
y m NA
M
ð2Þ
where n is the number of 115In, y is isotopic abundance of 115In, the
indium nuclide (95.7%), NA is Avogadro’s number (6.022 1023 in
atoms/g-atom), M is molar mass of the foil material and m is indium
foil mass.
The measured g-ray spectra were collected using a multichannel analyzer (10 þ ; Canberra Industries, Meriden, CT, USA).
These foils were immediately placed on the detector face. The e at
the characteristic g-ray energy of 417 keV was emitted from the
isomeric transition of 116m1In and calibrated as 4.0%. The efficiency
of the GC3520 HPGe detector measured a series of g-energy using
calibrated standards for 60Co, 133Ba, 137Cs and 152Eu. The e in counts/
g was extrapolated for a similar geometry. For each irradiation
position, the induced radioactivities of 116m1In from 15 and 10 MV
LINACs for count times of 3 and 6 min, respectively, were precisely
determined. Following exposure, the activated indium foils were
allowed to decay for appropriate times. The principal errors are
systemic and consist of detector energy responses. Spectra acquired
were identified and calculated using Micro SAMPO-90 software and
a personal computer. Details of the experimental conditions are
given elsewhere (Chen et al., 2007; Uwamino et al., 1986; Gur et al.,
1978). Statistical errors were within 10%.
Structure shielding design of maze and entrance slide door
was built to fulfill the recommendation of NCRP 49 (NCRP, 1976).
A maze and shielded entrance slide door is utilized to protect the
personnel outside the facility (Baker and Thomas, 2001). Point A
( 3.8, 5.9, 0) is on the maze centerline and visible from the
isocenter (Fig. 1). All measurements were performed by placing
indium foils at locations A and B, (7.8, 5.9 and 0; near the 10 cm
PE sliding door) at 0.5–1 m intervals along a straight line.
3. Results and discussion
3.1. The Fth distributions at 15 and 10 MV
Fig. 3(a) and (b) shows the foil technique results for Fth at 15 and
10 MV in the 21EX Varian clinic at CSMUH; Fth is a function of
distance from the central axis. The difference in Fth between 15 and
10 MV beams is attributed to the difference in X-ray energy of the
LINACs. Three-dimensional distributions of the Fth for the treatment
room can be mapped using colored profiles that reflect various
intensities of the thermal neutron field. The blue color in
Fig. 3(a) reflects the great Fth value 41.78 105 n cm 2/Gy-X and
3.6070.16 105 n cm 2/Gy-X was measured at the gantry head,
roughly 1.83 times greater than that at the isocenter on the z-axis.
This finding is expected as the contribution of scattered components
to the total Fth that decreases as the distance to the isocenter
increases as shown in (Fig. 3(a) and (b)). Both minor and significant
variations of Fth were observed across the treatment room. The
representative Fth values on the y-axis from the isocenter edge at 60,
100 and 160 cm are (1.3270.12) 105, (1.2670.11) 105 and
(1.2870.12) 105 n cm 2/Gy-X, respectively (Fig. 3(a)). The measured Fth, ranging from 1.78 105 to 1.57 105 (Fig. 3(a)) is shown
in red color; this finding is due to the distributions of leaking
neutrons through the accelerator head. The measured Fth around
the maze entrance at the far end of the isocenter (green color in
Fig. 3(a)) is (1.0170.09) 105 n cm 2/Gy-X; the lowest Fth in the
treatment room is o(1.0170.09) 105 n cm 2/Gy-X (yellow color
in Fig. 3(a)). The Fth at 15 MV was (1.9770.13) 105 n cm 2/Gy-X,
approximately 13.5 times greater than that of the 10 MV beam
((1.4670.13) 104 n cm 2/Gy-X) at the isocenter. The Fth in the
treatment room irradiated by 10 MV X-ray vary, (1.4670.13)
104 (6.3570.03) 103 n cm 2/Gy-X and are nearly uniform
throughout the treatment room (Fig. 3(b)).
3.2. Distributions of Fth in a water phantom
2.3. Field application for water phantom
The Fth measurements performed using the foil technique
simulated total body irradiation is carried out using a water
phantom from Varian CLINAC 21EX accelerators operating at
In this study, dose depth equivalent measurements were preformed in a water phantom irradiated with an X-ray beam field of
20 20 cm2 from Varian 21EX accelerators operating at 15 MV.
The photoneutrons entered the water moderator and slowed via
Author's personal copy
920
2×105
water level
2.00×105
1.78×105
1.57×105
1.35×105
1.13×105
9.10×104
6.90×104
1.0×105
1×105
0.5×104
-2
X
ax
is
2
)m
0
0
,y,
0
(
xis
0
(x
,0,
0)
2
-2
m
Ya
Thermal neutron fluence (n⋅cm-2/Gy-X)
Thermal neutron fluence
(n⋅cm-2/Gy-X)
W.-S. Liu et al. / Radiation Physics and Chemistry 80 (2011) 917–922
isocenter
Gantry head
z
y
water level
isocenter
105
tank bottom
1.50×104
1.32×104
1.13×104
9.40×103
7.51×103
5.65×103
0.4
0.5×103
0
X
-2
ax
is
(
0
,0)
2
)m
0
,
0,y
0
x,0
m
-2
2
xis
(
Ya
Fig. 3. Distributions of Fth in treatment room (Varian CLINAC 21EX) irradiated by
a 20 20 cm2 beams of (a) 15 and (b) 10 MV X-rays. Experimental data are
displayed using a three-dimensional colored graphical representation (thermal
neutron fluence unit is n cm 2/Gy-X). (For interpretation of the references to
colour in this figure, the reader is referred to the web version of this article.).
a
103
0.2
0.1
Z axis (0, 0, z) m
0.0
-0.1
-0.2
isocenter
105
CSMUH
isocenter
1m
x
104
-0.3
104
0.3
106
Thermal neutron fluence (n⋅cm-2/Gy-X)
utron fluence
Thermal ne-2
(n⋅cm /Gy-X)
0.5
1.5×104
1.0×104
tank bottom
106
-0.2
-0.1
0.0
0.1
X axis (x,0,0)
0.2
0.3
Fig. 5. Distributions of Fth in a water phantom irradiated by a 20 20 cm2 beam
of a 15 MV X-ray. (a) versus depth along the central z-axis; and (b) across the
width at a depth of 15 cm on the x-axis.
1293 keV 116m1In
101
2212 keV 23Th
460 keV 40K
104
417 keV 116m1In
100
336 keV 115mIn
Counts/Channel
102
102
103
101
100
b
1000
2000
3000
4000
The major and relatively significant variations of Fth, varying
from (8.8070.09) 105 (6.8570.07) 104 n cm 2/Gy-X, were
observed on the z-axis in the water phantom (Fig. 5(a)). The Fth
was a function of distance from the z-axis and penetration depth in
the water phantom. The depth of maximum Fth was 1.3 cm on the
z-axis that is, (0, 0, and 0.13). Moreover, the Fth was relatively
constant inside and outside the 20 20 cm2 field size along the
x-axis and roughly uniform throughout the water phantom
(Fig. 5(b)). At a depth of 25 cm (0, 0, and 0.10) from water
surface, Fth was only 31.1% of the isocenter of the water phantom,
the smallest among all water phantoms (Lennox, 2001).
3.3. Thermal neutron dose to patients
Channel
Fig. 4. Gamma-ray spectra collected by the germanium detection system for
radioactivity measurement of an indium foil of 20 20 cm2 exposed to 15 MV
X-rays. (a) isocenter and (b) (0.12, 0, 0) outside the beam field.
elastic collisions and captured to form 116m1In. The spectra in
Fig. 4(a) and (b) were measured at the isocenter and outside the
beam field (0.12, 0 and 0) of water phantom, respectively. The
photon peak at 336 keV is clearly seen in the spectra (Fig. 4(a)),
which is from 115mIn and is associated primarily with high-energy
photon interaction through 115In(g, g)115mIn (Chao et al., 2001).
Thermal neutron exposure is not considered in routine dosimetry planning; however, patients may require ensuring their
safety. McCall (McCall, 1997) stated that thermal neutron dose
conversion factor is 260 n/cm2 s¼ 1 mrem/h¼2.78 nSv/s. Therefore, based on experimental data obtained from the isocenter of
the water phantom, the thermal neutron dose equivalent is
9.42 mSv/min for a patient receiving a 10 Gy photon dose during
a routine 2.5 min treatment. For a 15 MV 20 20 cm2 field
size, the thermal neutron dose equivalent at the isocenter of the
water phantom is 2.36 mSv/Gy-X, representing a total dose of
Author's personal copy
921
W.-S. Liu et al. / Radiation Physics and Chemistry 80 (2011) 917–922
3.4. Fth in the maze
The Fth were measured along the line A–B (cf Fig. 1). The Fth at
point A was equal to (5.0170.03) 104 ncm 2/Gy-X, 25.4% of that at
the isocenter, which was higher than that at other places in the maze
due to neutrons leaking through the accelerator head (Comsan, 1996;
Kamino, 1998). The Fth of point B, (2.8570.21) 103 n cm 2/Gy-X
was approximately 1.45% of that at the isocenter and the smallest in
the treatment room. This finding states that photoneutrons are
emitted strongly forward after being produced by high-energy X-rays
from beam-modified devices. Fig. 6 shows a semi-log plot of the Fth
as function of distance (x) along the A–B centerline.
The Fth, x (Fig. 6) can be expressed as an exponential function
of x
Fth,x ¼ Fth 10TVL x
This work (Varian Clinac 21Ex)
Barquero et al., 2005 (Siemens)
Konelfal et al., 2005 (Varian, Siemens)
Lin et al., 2001 (Primus Siemens)
Uwamino et al., 1986 (Microtron MM22)
Plata et al., 1985(Mevatron 77)
Gur et al., 1978 (Philips, SL/75-20)
Deye and Young 1977 (Varian Clinac)
McGinley, 1976 (Varian Clinac-18)
3×106
2.5×106
2×106
1.5×106
1×106
5×105
0
5
10
15
20
25
30
Accelerator energy (MV)
Fig. 7. The neutron fluence at the isocenter of medical accelerator operated at
different peak X-ray energies.
ð3Þ
where Fth, x is the Fth of the x position, TVL is tenth-value layer
(m) of Fth in the maze. The TVL can be obtained from Fig. 6 by
extrapolating the portion of curve with a y value to a point on the
A–B straight line. The TVL was estimated at 7.2770.10 m in this
study. This experimental result is 4 6.4 m, as obtained by
McGinly and Miner, who proposed that a TVL 46.4 m can be
classified as hard or the high-energy component of neutrons
(McGinley and Miner, 1995). At point ‘‘S’’ outside the sliding
door, the Fth were negligible or below the detection limit (DL) of
the indium foil activation approach, indicating that 10 cm PE
effectively attenuates the penetrating neutrons and an adequate
shielding design of the wall and maze (cf. Fig. 1).
3.5. Detection limit (DL) of foil activation technique
The DL of Fth for 116m1In in this study was 4.65 DS (B/t)1/2,
where B is the background count rate at the 417 keV of 116m1In,
and t is the counting period (Chen, 2003; Currie, 1968). Hence, the
detection sensitivity (DS) for 116m1In can be calculated by dividing
the counts of 417 keV by the e and absolute g-ray intensity. In the
6 min counting period of the 10 MV LINAC, the DL of Fth was
Thermal neutron fluence (n⋅cm-2/G-X)
3.5×106
Neutron fluence (n⋅cm-2/Gy-X)
2.36 10 4%, which is in agreement with the experimental
results reported elsewhere (Lin et al., 2001; Paredes et al.,
1999). Additionally, the thermal neutron equivalent near the
treatment bed is roughly 10 5 of the dose delivered to patients.
estimated at 400 n cm 2/Gy-X (Currie, 1968). Using this technique to determine Fth in treatment rooms for LINAC at hospitals is
feasible.
3.6. Fth compared with results obtained by other studies
Fig. 7 shows the neutron fluence measured at the isocenter
within a 20 20 cm2 X-ray field and expressed in n cm 2/Gy-X.
The Fth obtained are compared with those obtained by recent
studies. The only exceptions relative to Fth are experimental
results acquired by Lin et al. (Lin et al., 2001), who used a
40 40 cm2 field size. Generally, these studies suggested that
the range of Fth was 1.20 104 1.60 106 n cm 2/Gy-X at
100 cm SSD for LINACs operating at 10–25 MV. Experimental data
in this study are in agreement with the average of results
obtained by other studies (Barquero et al., 2005; Konefal et al.,
2005; Lin et al., 2001; Palta et al., 1984; Gur et al., 1978; Deye and
Young, 1977; McGinley et al., 1976). This agreement is demonstrated by the linear scale for Fth (Fig. 7). Conversely, the Fth for
15-MX Varian CLINAC 21EX is roughly 26% lower than that of
Primus Siemens (Lin et al., 2001). This difference is due to the
different materials used to construct the accelerator head of the
target collimators and scattering shields.
105
4. Conclusion
A
B
104
HPGe GC3520
10
10+MCA
PC / SAMPO-90
3
0
2
4
6
8
10
Maze (m)
Fig. 6. Exponential fitting curves of Fth in the maze. Locations of the Fth
estimation (A–B) points are shown with the corresponding distances from the
maze entrance.
The Fth produced from a Varian CLINAC 21EX operated at 15 and
10 MV were successfully measured and the accompanied neutron
dose for personnel was also elucidated herein. The foil activation
technique is an effective method of measuring Fth in treatment
rooms. The Fth for the 15 MV, (1.9770.13) 105 n cm 2/Gy-X was
roughly 13.5 times that of the 10 MV beam, (1.4670.13)
104 n cm 2/Gy-X at the isocenter. The Fth measurements for Varian
CLINAC 21EX accelerators indicate that Fth is a function of distance
from the central axes of x and y and penetrating depth in the water
phantom, representing only 2.36 10 4% of the total dose. The Fth
of the Varian 21EX are in agreement with those reported in literature
(McGinley and Miner, 1995). Thus, a 10 cm thick PE is capable of
suppressing the neutron dose for personnel and is also a reliable way
to reduce the Fth at the outer maze entrance. A maze with 10 cm
thick PE effectively attenuates the penetrating neutrons and provides
adequate shielding for the wall and maze.
Author's personal copy
922
W.-S. Liu et al. / Radiation Physics and Chemistry 80 (2011) 917–922
Acknowledgements
The authors wish to thank radiotherapists of CSMUH for their
assistances in the operation of the accelerator. This work was
financially supported by the National Science Council of Republic
of China under Contract no. NSC 97-2221-E-040-008.
References
American Association of Physicists in Medicine, 1986. Neutron Measurements
Around High Energy X-ray Radiotherapy Machine. AAPM No. 19, New York.
Baker, C.R., Thomas, S.J., 2001. Neutron transport in a clinical linear accelerator
bunker: comparison of materials for reducing the photo-neutron dose at the
maze entrance. Radiat. Phys. Chem. 61, 633–634.
Barquero, R., Mendez, R., Vega-Carrillo, H.R., Iniguez, M.P., Edwards, T.M., 2005.
Neutron spectra and dosimetric features around an 18 MV LINAC accelerator.
Health Phys. 88 (1), 48–58.
Chao, J.H., Hsu, P.C., Liu, H.M., 2001. Measurement of high dose rates by photon
activation of indium foils. Appl. Radiat. Isot. 55, 549–556.
Chen, C.Y., 2003. Optimal conditions for identifying 80Br and 128I in health food
Angelica keiskei using rapid epithermal neutron activation analysis. Appl.
Radiat. Isot. 58 (4), 423–429.
Chen, C.Y., Lin, D.B., Chen, W.K., 2007. Elements in the sera of preschool children
living in central Taiwan. Chin. J. Chem. 25, 515–520.
Comsan, M.N.H., 1996. Shielding calculations for Inshas cyclotron. Radiat. Phys.
Chem. 47 (5), 735–738.
Currie, L.A., 1968. Limits for qualitative detection and quantitative determination.
Anal. Chem. 40, 586–593.
Deye, J.A., Young, F.C., 1977. Neutron production from 10 MV medical LINAC. Phys.
Med. Biol. 22 (1), 90–94.
Gur, D., Rosen, J.C., Bukovitz, A.G., Gill, A.W., 1978. Fast and slow neutrons in an
18 MV photo beam from a Philips SL/75-20 linear accelerator. Med. Phys. 5 (3),
221–222.
Kamino, Y., 1998. 10 MeV 25 kW industrial electron LINAC. Radiat. Phys. Chem. 52
(1–6), 573–576.
Kase, K.R., Mao, X.S., Nelson, W.R., Liu, J.C., Kleck, J.H., Elsalim, M., 1998. Neutron
fluence and energy spectra around the varian CLINAC 2100C/2300C medical
accelerator. Health Phys. 74, 38–47.
Konefal, A., Dybek, M., Zipper, W., Łobodziec, W., Szczucka, K., 2005. Thermal and
epithermal neutrons in the vicinity of the Primus Siemens biomedical accelerator. Nukleonika 50 (2), 73–81.
Lennox, A.J., 2001. Accelerators for cancer therapy. Radiat. Phys. Chem. 61,
223–226.
Lin, J.P., Chua, T.C., Lin, S.Y., Liu, M.T., 2001. The measurement of photoneutrons in
the vicinity of a Siemens Primus linear accelerator. Appl. Radiat. Isot. 55,
315–321.
McCall, R.C., 1997. Shielding for thermal neutrons. Med. Phys. 24 (1), 135–136.
McGinley, P.H., Miner, M.A., 1995. Method of eliminating the maze door of medical
accelerator rooms. Radiat. Prot. Manage. 12 (5), 29–37.
McGinley, P.H., Wood, M., Millis, M., Rodriguez, R., 1976. Dose levels due to
neutrons in the vicinity of high-energy medical accelerators. Med. Phys. 3 (6),
397–402.
National Council on Radiation Protection and Measurements, 1976. National
Council on Radiation Protection and Measurements Report 49. Structural
Shielding Design and Evaluation for Medical Use of X-rays and Gamma-rays
of Energies up to 10 MeV, Washington, DC.
Palta, J.R., Hogstrom, K.R., Tannanonta, C., 1984. Neutron leakage measurements
from a medical accelerator. Med. Phys. 11 (4), 498–501.
Paredes, L., Genis, R., Balcazar, M., Tavera, L., Camacho, E., 1999. Fast neutron
leakage in 18 MeV medical electron accelerator. Radiat. Meas. 31, 475–478.
Uwamino, Y., Nakamura, T., Ohkubo, T., Hara, A., 1986. Measurement and
calculation of neutron leakage from a medical electron accelerator. Med. Phys.
13 (3), 374–384.